GEA Niro Method No. A 14 a |
|
|
Revised: September
2005 |
|
1.
Definition
The hygroscopicity of a powder is its
equilibrium moisture content after being exposed to air humidity under given
conditions.
2.
Scope
This method is particularly suitable for
whey powder, but may be applied for other dried milk products.
3.
Principle
Air with a 79.5% relative humidity (RH) is
sucked through the powder sample until a constant increase in weight has been
reached.
4.
Apparatus
1.
Analytical balance, sensitivity ±0.1 mg
2. Vacuum
flask, 500 ml.
3. Gooch
filter with ground cone 34/35, porosity 1.
4. Gooch
filter adapter with ground glass socket 34/35, see Fig. 1.
5. Glass
covering tube assembled by means of 2 rubber stoppers to the vacuum flask, see
Fig. 1.
6. Water
vacuum pump.
7. Flow
meter, 0-400 ml/min.
8. Washing
bottle, 250 ml.
9. Three way
glass cock.
10.
500m sieve.
5.
Reagents
Ammonium chloride, saturated solution at 20°
C.
6.
Procedure
1. Assemble
the apparatus with rubber tubes as follows (see Fig. 1):
1.1 vacuum
pump
1.2 three-way cock
1.3 vacuum flask with
filter and filter adapter
1.4 washing
bottle
1.5 flow
meter
One position of the three-way cock should allow passage of air through
the apparatus. The other passage of air directly from the atmosphere to the
pump and at the same time keeping the apparatus tightly closed from the
atmosphere.
2. Fill the
washing bottle with a saturated solution of ammonium chloride and add a surplus
of ammonium chloride crystals.
3. Place a
clean and dry Gooch filter in the adapter and assemble as shown in Fig. 1. Turn
the cock to position 1 (pump atmosphere). Start the vacuum pump. Turn the cock
to position 2 (pump washing bottle), set the flow rate to 3-400 ml/min and pump
for 5 minutes.
4. Turn the
cock to position 1 and wait until the circulation through the washing bottle
has ceased. Open the apparatus and weigh the Gooch filter.
5. Weigh out
approx. 0.5 g of the sample on the filter with an accuracy of 0.1 mg of powder,
which has been forced through a 500 m sieve. Spread the
powder evenly over the filter bottom by tapping gently.
6. Place the
Gooch filter in the apparatus and start as described in
6.3.
7. Check the
flow rate from time to time and clean the washing bottle if necessary (the tube
can be blocked by crystals).
8. Check the
increase in weight every 10 min. during the first 30-40 min. and then every 20
min. Stop the air flow as described in 6.4.
The increase in weight
normally shows a maximum and then slightly decreases and reaches a steady
level. When the maximum has been reached the analysis can be stopped. Normally
this does not take more than 4 hours. If degree of caking (Method
A 15 a) has to be determined, keep the sample in the apparatus until
it can be transferred to the oven.
7. Result
Calculate the result to 1 decimal
8.
Reproducibility
±
1% relative
9.
Remarks
1. All whey
powders are more or less hygroscopic, and it is therefore important to
determine free water content at the same time as the
hygroscopicity.
|
Hygroscopicity |
|
Non hygroscopic:
Slightly hygroscopic:
Hygroscopic:
Very hygroscopic:
Extremely hygroscopic: |
 
; ; <10%
10.1-15%
15.1-20%
20.1-25%
>25% |
 
;
10.
Reference
GEA
Niro Research Laboratory